VA-Clave

96×96

4 Digit – Dual Row
Auto Clave Controller

Features

  • RTD : Pt-100 Temperature Input
  • Pressure Input – 0~10VDC, 4~20mA
  • Input Signal Conditioning (Digital Filter & Zero Offset)
  • F0 calculation at the end of Sterilization
  • Drying facility to avoid high outrush of pressure
  • Sterilization Timer with Guaranteed Soak Time
  • High Temperature Safety Alarm
  • Automatic Cycle Abortion at High Temperature Limit
  • Independent Air Purging and Sterilization Setpoint
  • Independent Outputs for Air Outlet & Steam Control
  • Front / Remote Sterilization Cycle Start Command
  • Optional Serial Communication Port
  • Water Level Switch Input
  • Universal supply voltage from 90 to 270 VAC
  • Supply 24 VDC Available on Request
DescriptionVA-Clave
DisplayHeight & Colour0.80'' White & 0.56'' Green
No. of Digits4 Digits Dual Row
LED IndicationHTR, P/E, ALM, ST, SP, PR
Input Specification
InputTemperature InputRTD: Pt-100
Pressure Input0-10 VDC, 4-20 mA
ResolutionTemperature : Pt-100: 0.1°C
Pressure : 1/ 0.1/ 0.01/ 0.001
Sampling Rate250msec
Accuracy0.3% FS
Excitation Voltage5V/ 10V/ 12V/ 24V (Jumper Selectable)
Serial Communication
PortRS 485, Half Duplex
ProtocolModbus RTU
Baud Rate9600, 19200, 31250, 38400, 76800
ParityNone, Odd, Even
Data Bits8
Stop BitsOne(1), Two(2)
Output Specification
Output Relay5A @ 250 VAC/ 30VDC
SSR Drive12 VDC @ 30mA
General Specification
Supply Voltage
90 - 270 VAC/ DC
(24 VDC Available on Request)
Temperature And HumidityOperating: 0-50°C
Storage: -20° to 75°C
95% RH (Non-condensing)
Dimensions in mm (H x W x D)96 x 96 x 65